KASON HTBRV-i250 intelligent fully automatic Brinell, Rockwell and Vickers Hardness Tester
Product Introduction
1. Modern minimalist design.
2. A variety of Brinell, Rockwell and Vickers scales, HVT and HBT are available
3. Brinell and Vickers measurement objectives and light sources are separate and independent;
4. Brinell and Rockwell pressure Test completed, automatically move to the measurement position, and automatically measure, one click to obtain hardness value;
5. Fully automatic, high-precision force sensor, closed-loop force feedback control system
6. Workbench automatic lifting, anti-collision detection device and Z-axis retraction system
7. Automatic application of initial test force and primary test force.
8. The worktable automatically rises, stops rising when the sample touches the indenter, automatically loads, automatically unloads, and automatically measures the hardness value.
9. Touchscreen interface, easy to operate, directly select the hardness scale, force value changes automatically.
10. Powerful data processing capabilities, including: conversion of multiple hardness systems such as HR, HB, HV, qualified range setting, automatic alarm for over-limit, and input of Tester, sample name, etc
11. According to the standard hardness blocks, each hardness scale is automatically corrected in three segments: high, medium and low.
12. The test force can be automatically corrected by the standard dynamometer.
13. Ultra-high precision electric X/Y worktable
Repeat positioning accuracy: Up to 200mm for a 3-micron travel
14. The database processing function is powerful, including: conversion of multiple hardness systems such as HR, HB, and HV, setting of qualified range, automatic alarm when exceeding the limit, and input of information such as tester and sample name. Measurement parameters and data are automatically saved to the database.
15. Qr code scanning, cloud storage of detection data, data analysis, remote diagnosis and upgrade.Real-time display of immediate test data and historical data on mobile phones and computers.
16. Automatic calculation of hardness value uncertainty (option)
17. Automatic generation of CNAS review required Documentation (Option)
18. Remote Diagnosis and Upgrade (Option)
Technical parameters:
Model |
KASON HTBRV-i250 |
Hardness scale |
Brinell, Vickers, Rockwell, surface Rockwell |
Force value |
5 to 250kgf |
Brinell scale |
HBW1/ : 5, 10, 30kgf HBW2.5/ : 6.25, 15.625, 31.25, 62.5, 187.5kgf HBW5/ : 25, 62.5, 125, 250 KGF HBW10/ : 100, 250kgf |
Rockwell scale |
A, B, C, D, E, F, G, H, K, L, M, P, R, S, V 15N, 30N, 45N, 15T, 30T, 45T, 15W, 30W, 45W 15X, 30X, 45X, 15Y, 30Y, 45Y |
Vickers scale |
HV5, HV10, HV20, HV30, HV50, HV100, HV120kgf |
Measurement system |
Automatic measurement |
Objective magnification |
2.5X, 5X (optional 10X, 20X) |
Camera |
500 pixels |
Data shows |
Computer shows data |
Test |
Auto-load, hold, unload |
Press head |
Rockwell diamond: 120° Rockwell ball indenter: 1/16" Vickers diamond: 136° Brinell ball indenter: 1mm, 2.5mm, 5mm, 10mm (Rockwell ball press options 1/8 ", 1/4 ", 1/2 ") |
Illumination system |
LED |
enforce standards |
GB/T230, GB/T231, GB/T4340, JJG144-1999, GB/T18449, ISO 6508, ASTM E10, ASTM E92, ASTM E18, ASTM E384, ASTM E103 |
Total test force holding time |
1-99 seconds |
Sample specifications |
Maximum height of sample: 220mm Distance between the indenter and the outer wall (throat) :200mm Cylindrical specimens: The minimum diameter of cylindrical specimens on the outer surface is 4mm; |
Maximum test height |
Rockwell: 220, Brins-Vickers: 150mm |
Distance from the center of the indenter to the outer wall |
200mm |
Size |
500*260*750mm |
Weight |
About 70Kg |
Power supply |
AC220V+5%, 50-60Hz |